GCE Products for COVID-19 CHECK IT OUT close
Home » EASE II

EASE II

Group Number: 3521

Whole medical range is available on

EASE II demand valve is a robust and compact device used by patients to self administer medical gas therapy. EASE II can be used to administer nitrous oxide and oxygen mixture (commonly known as N2O/O2) for pain relief or medical oxygen therapy. The EASE II demand valve is designed in a way that creates minimal breathing resistance to the patient and can deliver high flows when required.

EASE II O2 is recommended during diving accidents and cluster headaches therapy.
EASE II N2O/O2 is recommended during pain releave therapy.

FEATURES

  • Deliver up to 300 litres per minute of gas
  • Portable first stage regulator and cylinder version for immediate care and pre-hospital applications
  • Conforms to global standards
  • Low inspiratory effort demand valve
  • Test/ Purge facility on the demand valve
  • Easy grip handle and wrist strap
  • Replaceable patient/bacterial filter
  • Easy to clean and reassemble for cross infection protocol
  • Hose fitted with probes by the national standards for connection into cylinder system or wall outlet
  • Autoclavable at 134 °C

 735100000436 EASE II leaflet EN


ACCESSORIES

Art. Nr. Description
1032936 Disposable mouthpiece
1035575 Breathing filter
8848004 Reusable universal size mask
1024417 Blue barrel carry bag
5487294 Mask adult size single use
1032994 Mask adult size single use - vanilla
850500 Expiration diverter single use for Ease II
1570000SP Scavenging hose

TECHNICAL DATA

DEMAND VALVE

Gas: N2O/O2
Material: Polycarbonate, silicone rubber, stainless steel
Dimensiones: 50 × 50 × 63 mm
Weight: 0,085 kg
Gas supply: Requirement 2,8 to 7,0 bar at > 200 l/min

Inspiratory resistance
at 2,8 bar supply press.:

-0,15 to -0,2 kPa at 0 l/min

-0,2 kPa at 10 l/min

-0,7 kPa at 200 l/min

Expiratory resistance: +0,35 kPa at 120 l/min
Operating temperature: -20°C to +60°C used Oxygen

+5°C to +40°C used 50/50 N2O/O2
Storage temperature: -30°C to +60°C
Regulatory status:

Complies with Medical Devices Directive 93/42/EEC

Complies with BS 4272: Part 2: 1996

Production in accordance with EN ISO 9001 and EN ISO 13485

 HOSE ASSEMBLY

Connection: Probes by the national standards
Working pressure: 7 bar
Burst pressure: 44 bar
Material: PVC, anti-static in accordance with ISO 5359
Weight: 0,5 kg (3 m length)
Manufacturer:

GCE, s.r.o, Žižkova 381, 583 01 Chotěboř, CZ